Specifications
Interpretive Stress Test withTreadmill
Patient cables digitizes the signal at the patient,
eliminating analog cable noise
Contacts resist corrosion and ensure optimal signal
Quality
Median update help s assure all measurements, including
ST segments, are made upon true morphology
Technician friendly interface. Select standard or
customized protocols.
Display
Size: 30.48 cm (48 in) diagonal.
Type: high-resolution, raster written.
Frequency response: 0.01 to 40 Hz, 0.01 to 20 Hz,user-selectable.
Number of traces: 3 rhythmleads.
Sweep speed: 25 mm/s.
Sweep speed accuracy: ±2%.
ECG rhythm: 7.5 s/lead.
Leads: user-selectable.
Recall reports: 10s of 6 user-selectable leads.
Displayed data: heart rate, target heart rate, blood pressure, total exercise time clock, stage clock, 4 current median
complexes, treadmill speed and grade, ventricular ectopic/minute counter, 3 channels of rhythm, protocol and stage of exercise, warning messages and
prompts.
System
Digital sampling rate: 250 Hz. A/D conversion: 12 bits. Chassis leak age current: <100 mA(meets AAM I standards SCL 1278).
Patient leakage current: <10 mA. Common mode rejection: 130 dB, minimum, up to 180 dB with digital processing. AC elimination: 50/60 Hz digital filters.
Electrode offset tolerance: ±320 mVDC. Internal maintenance test: automatic self-tests. Diskette drive: 8.89 cm (3.5in), 2 MB, double sided,
quad-density, and MS-DOS format compatible.
Processing
ST level and slope determination: up to 15 leads. Medians: incremental updating with elimination of artifact and aberrance in all leads. ST level
measurement: manually selectable post J point. Heart rate: 16-beat sliding average updated every beat; minimum 35 beats/min, maximum 300
beats/min. QRS detection: band-pass filtered composite of three user-selectable leads. Arrhythmia: automatic detection and documentation.
Input Channels: simultaneous acquisition of up to 15 channels. Impendence: 100 M1/2, defibrillator-protected. Dynamic range: ±320 mVDC.
Electrode offset tolerance: ±320 mVDC.
Output
Standard outputs: one analog output , one wave form output, one TTL output with variable pulse width, polarity and delay, QRS beep
Writer Method: therma l dot array (horizontal 1000 dots/in at 25 mm/s, vertical 200 dots/in). Channels: 6, user selectable. Frequency response: 0.01 to 100
Hz, 0.01 to 40 Hz, 0.01 to 20 Hz, user-selectable. Sensitivity: 5,10, and 20 mm/mV. Paper speed: 5,12.5, 25,and 50 mm/s, user-selectable.
Paper style: thermal, perforated, 21.59 cmx 27.94 cm x (8.55″ x 11″), Zfold, 300 sheets/pack.
Acquisition Module
Analog Sampling rate: 2000 Hz . Microprocessor-controlled: 10 IEC and AHA color-coded lead wires standard, 14 lead wires optional; (additional
lead wires user-definable)
Keypads
Control panel: contains manual treadmill, test/procedures, and report controls. Alphanumeric keyboard: with TRIM KNOB for blood pressure, patient demographics, and impressions entry
Reports
Classical 12-lead: note: reports can be transferred to the NUSE network by diskette. Incrementally averaged medians: with real-time rhythm and ST
level; user-selectable in two formats. 3- or 6-channel rhythm strips: available for real-time, recall or arrhy thmia documentation (leads user-selectable).
Tabular presentation of final report data including: phase, stage, speed, grade, heart rate, blood pressure, rate-pressure product, ectopic, maximum ST
depression, and physician’s impressions. Sample cardiac cycles: presented at user-defined intervals. Trend graph: heart rate, ST levels, ST slopes,
blood pressure, PVCs/m.
Power
Requirements: 90 to 135 VAC, 180 to 270 VAC, 50/60 Hz. Power consumption: 225 W maximum.
Physical
Height: 154.94 cm (61 in).Width: 62.23 cm (24.5 in). Depth: 71.12 cm (28 in). Weight: 54.48 kg (120 lbs)