Stress Test Electrodes
Gel Monitoring Electrode
1 1/2″ Foam, Wet Gel Monitoring Electrode
Duration of Use (Hrs): 24 |
Application: Stress and Holter Testing |
PRODUCT INFORMATION | |
Shape: Teardrop |
|
Size (excl. grip): | |
1.5 in | 38.1 mm |
Sensor (Eyelet) Diameter: | |
0.42 in | 10.592 mm |
Substrate Thickness (adapter excluded): | |
0.063 in | 1.588 mm |
Total Product Surface Area: | |
1.7972 in2 | 45.649 mm2 |
Gel Area: | |
0.3068 in2 | 7.7927 mm2 |
Adhesive Area: | |
1.4904 in2 | 37.8562 mm2 |
Integrated Lead Wire (length): N/A in |
|
Integrated Lead Wire (color): N/A |
MATERIALS |
Substrate Material: Polyethylene Foam Adhesive: Acrylic, Medical Grade Adhesive Strength: High Tack Gel Type: Conductive Wet Gel Foam (Sponge) Material: Polyurethane Reticulated Foam Release Liner: PS Sensor Polymer: Ag/AgCl plated ABS Adaptor / Connector (Stud): Stainless Steel Integrated Lead Wire Jacketing: N/A Integrated Lead Wire Cord: N/A |
ELECTRICAL PERFORMANCE (ANSI / AAMI EC12) |
|
ACZ impedance @10 Hz: 32 Ohm DC Offset Voltage : 0.38 mV SDR Slope @30 Sec int.: 0.07 mV/sec |
ACZ impedance repeat: 25 Ohm COIIN: 7 µV Bias Current Tolerance: 7 mV |
FEATURES | |
MRI Conditional: No X-ray Translucence: No |
Integrated Abrader: No Repositionability: No Pediatric: No |
PACKAGING | |
Product Packaging Material: | |
Paper/PE/Foil/PE | |
Resealable Pouch: | |
No | |
Product Packaging Size (L x W): | |
11.5 x 3.75 in | 29.21 x 9.525 cm |
BIOCOMPATIBILITY | |
ISO 10993: Passed |
Latex Free: Yes |
ENVIRONMENTAL | |
Halogenated Hydrocarbon Content (e.g. PVC): None Phthalate Derivatives Content (e.g. DEHP): None |
RoHS Compliant: Yes REACH Compliant: Yes |
SHELF LIFE |
Product Shelf Life (in accordance with storage guidelines): 18 Months |
REGULATORY |
Regulatory Status: CE Mark according to MDD 93/42/EEC CE marked |
Use with your GE,Marquette, Quinton, Burdick, Mortara, Schiller and Philips stress treadmill system.