Physio-Control Lifepak 20 Specifications
Dimensions
Height: 8.4″ (21.3 cm)
Width: 10.3″ (26.2 cm)
Depth: 10.3″ 26.2 cm)
Weight: fully featured defibrillator/monitor (pacing and SpO2) 5.58 kg (12.3 lbs)
Power
The device is an AC line-operated device with an internal battery as backup.
AC Power: 90-132 VAC 50/60 Hz, 198-264 VAC 50/60 Hz
total power draw less than 75 Watts
Internal battery backup: NiMh, batteries charge while the device operates from AC power
Operating time: a new fully charged internal backup battery will provide the following prior to shutdown: 120 minutes of monitoring, 135 minutes of monitoring without pulse oximeter, 90 360 joules discharges, and 70 minutes of monitoring plus pacing
Typical battery charge time: <2 hours when the device is powered off and AC power is applied
Low battery indication and message: when the device is unplugged from AC power, it switches to the battery. When the battery gets low, the battery detection icon is indicated with a low battery message in the status area, and a warning tone occurs
Display
Height: 3.4″ (8.63 cm)
Width: 4.53″ (11.5 cm)
Resolution: 230 x 240 dot color active LCD
Display Time: a minimum of 4 seconds of ECG and alphanumeric for values, device instructions, or prompts. Option to display one additional waveform.
Sweep Speed: 25 mm/sec for ECG.
Data Management
The device can easily print a CODE SUMMARY report, including an introduction to the patient information and critical event record. The summary report also includes event and vital signs log, and waveform associated with certain events.
Communications
The device can print archived patient records and has two data communications port, which supports a serial data cable.
Monitor
ECG: can be monitored through 3-wire or 5-wire ECG cables. Standard paddles or therapy electrodes are used for paddle lead monitoring
Lead selections: Leads I, II, and III with a 3-wire ECG cable, leads I, II, III, AVR, AVL, and AVF, V acquired simultaneously with a 5-wire ECG cable
ECG size: 4, 3, 2.5, 2, 1.5, 1, 0.5, 0.25 cm/mV
Heart rate display: 20-350 bpm digital display
Voice prompts: used for selected warnings and alarms, configured on/off
SpO2: saturation range: 1 to 100%
Saturation accuracy: (70-100%) (0-69% unspecified): Adults/ Pediatric: ±2 digits with no motion, ±3 digits with motion. Neonates: ±3 digits with or without motion
SpO2 update averaging rate: user selectable 4, 8, 12 or 16 seconds
SpO2 measurements: functional SpO2 values are displayed and stored
Pulse rate range: 25 to 240 pulses per minute
Pulse rate accuracy: ±3 digits without motion and ±5 digits with motion
Frequency Response
Diagnostic: 0.05 to 150 Hz or 0.05 to 40 Hz (user configurable)
Monitor: 0.67 to 40 Hz or 1 to 30 Hz (user configurable)
Paddles: 2.5 to 30 Hz
Analog ECG Output: 0.67 to 32 Hz (except 2.5 to 30 Hz for paddles ECG)